ASTM Type I Ultrapure Water.For example, for:
- Molecular biology, microbiology, PCR, inorganic and organic trace analysis
- HPLC, ICP, TOC-analyses
ASTM Type II Ultrapure Water.For example, for:
- Rinsing laboratory glassware
- Preparing and diluting buffers, reagents, tissue culture media and dyes
- Sample preparation for analytical methods such as flame AAS
|
 |
| Instantly and Fresh. |
|
 |
With TKA Smart2Pure
This is the password for converting normal tap water to ultrapure water of ASTM Types I and II.
All in one! TKA Smart2Pure is a compact complete-system that produces 3 or 6 litres of ultrapure water per hour. The built-in 6 litre tank for ultrapure water has a conical bottom for complete emptying. It is equipped with a sterile vent filter and a pressure pump for supplying equipment connected downstream (e.g. autoclaves, analyzers). If required, a flexible dispenser with a sterile filter can be connected to the tank. The system has a small footprint, is very attractively designed and looks good in any laboratory.
TKA Smart2Pure operates with the same high-tech purification procedures as the well known TKA Pacific and GenPure systems. |
High Performance Filter!
Option of an integrated ultrafilter, generously dimensioned for pyrogene-free and nuclease-free water. The automatic self-cleaning gives years of service. |
UV-Oxidation 185/254 nm!
Optionally with additional UV-oxidation for reducing the content of microorganisms and their metabolites. |
Powerful performance!
- Permeate flow 3 or 6 l/h
- Flow rate 1.0 l/min
- 6-Litre tank with integrated pressure and recirculation pumps
|
Lasting economy!
- Water purification progresses through separate modules, each with quick connects for simple individual replacement
- Module 1
Pretreatment module with integrated reverse osmosis membrane
- Module 2
Cartridge containing high-quality ultrapure resins with very long service intervals
|
Easiest to operate!
- Just press a button to get ultrapure water - no adjustments necessary, the microprocessor control controls itself
- Ultrapure water quality is ensured by the limiting value input
- RS 232 Interface
- The display can be tilted to your optimal reading position
|
Placement anywhere
- Stand it on the laboratory bench
- Mount it on the wall
|
Usable worldwide!
- Whether 220/230 V, 110/120 V or other voltages - and even with fluctuating voltages - the automatic voltage regulator continually supplies the system with a constant 24 volts
|
Almost maintenance-free!
- The tank is made of ultrapure water resistant, pigment-free and food-safe polyethylene and has a conical bottom outlet for complete emptying
- Can be easily cleaned and disinfected
|
| |
|
 |
| Specifications: |
|
 |
|
TKA Smart2Pure Standard |
TKA Smart2Pure UF* |
TKA Smart2Pure UV** |
TKA Smart2Pure UF/UV |
| Permeate performance at 15°C: |
3 l/h or 6 l/h |
3 l/h or 6 l/h |
3 l/h or 6 l/h |
3 l/h or 6 l/h |
| Flow rate: |
1.0 l/min |
0.6 l/min |
1.0 l/min |
0.6 l/min |
| Operating pressure in bar, min/max: |
1 – 6 bar |
1 - 6 bar |
1 - 6 bar |
1 - 6 bar |
ASTM I conductivity: Resistance at 25°C: TOC Value: Bacterial content in CFU/ml: Particle content: Endotoxines in EU/ml: |
0.055 µS/cm 18.2 MW×cm 5 - 10 ppb < 1 < 1
|
0.055 µS/cm 18.2 MW×cm 5 - 10 ppb < 1 < 1 0.005 |
0.055 µS/cm 18.2 MW×cm 1 – 5 ppb < 1 < 1
|
0.055 µS/cm 18.2 MW×cm 1 – 5 ppb < 1 < 1 0.005 |
ASTM II conductivity: Resistance at 25°C: |
0.067 – 0.1 µS/cm 15 – 10 MΩ×cm |
0.067 – 0.1 µS/cm 15 – 10 MΩ×cm |
0.067 – 0.1 µS/cm 15 – 10 MΩ×cm |
0.067 – 0.1 µS/cm 15 – 10 MΩ×cm |
| Supply voltage: |
Automatic voltage regulation to 24 V |
Automatic voltage regulation to 24 V |
| Power consumption: |
0.06 kW |
0.06 kW |
0.06 kW |
0.06 kW |
| Connector size: |
R 3/4" |
R 3/4" |
R 3/4" |
R 3/4" |
| Dimensions, W x D x H in mm: |
305 x 400 x 545 |
305 x 400 x 545 |
305 x 400 x 545 |
305 x 400 x 545 |
| Weight: |
22 kg |
22 kg |
22 kg |
22 kg |
| |
| Article no., 3 l/h version: |
08.2030 |
08.2031 |
08.2032 |
08.2033 |
| Article no., 6 l/h version: |
08.2060 |
08.2061 |
08.2062 |
08.2063 |
| * supplemented with an ultrafilter ** supplemented with UV-oxidation, 185/254 nm |
|
|
|
 |
| |
|
| |
|
| back >> |